BIOSEN

BIOSEN C-LINE. LABORATORY GLUCOSE AND LACTATE ANALYSER


Access all BIOSEN products in our online store: VER

The Biosen lactate analyser is considered an accurate and reliable laboratory analyser that can test blood, plasma or serum to provide glucose and lactate values with excellent accuracy (less than 2% CV) over a wide measurement range.

Biosen uses a special chip sensor technology in order to provide fast measurements with a high degree of accuracy at a low cost per test. Thanks to the long life of the chip sensor, the system requires little maintenance; features such as the touch screen and large memory only add to the ease of use of the Biosen analyser.

The Biosen analyser is available as a single-channel system, which allows one parameter to be analysed at a time, or as a dual-channel system that can measure lactate and glucose at the same time.

There are two Biosen models that differ in the number of samples that can be analysed simultaneously:

Biosen C-Line GP+ (up to 5 samples)
Biosen C-Line Clinic (up to 20 samples)

TECHNICAL SPECIFICATIONS:

Easy to use:

  • Multi-lingual touch screen display with step by step instructions
  • Just 20 μl blood, plasma or serum sample required when used with EKF pre-filled ‘Safe-Lock’ reaction cups (1,000 μl)
  • Up to 120 results an hour
  • Results in 20-45 seconds
  • Automatically detects presence of reaction cups
  • Chip sensors easily exchangeable, no electrode cleaning required

Accurate and reliable:

  • Enzymatic-amperometric method using chip-sensor technology
  • Measuring range: Glucose 0.5–50 mmol/L (9–900 mg/dL); Lactate 0.5–40 mmol/L (5–360 mg/dL)
  • Imprecision: CV ≤1.5 % (12 mmol/L)
  • Pre-filled ‘Safe-Lock’ reaction cups eliminate manual dilution
  • Innovative needle and exchanger design eliminates cross contamination
  • Three calibration modes
  • Método enzimático-amperométrico con tecnología de chip-sensor

Maximum efficiency and convenience:

  • Dual channel option: Glucose and/or lactate results from one sample
  • Long sensor life: Glucose 60 days / 7,500 tests; Lactate 50 days / 6,000 tests
  • Sample stable for up to five days (refrigeration recommended)
  • STAT position 

Efficient data handling:

  • PC, network and printer connectivity via RS232C interfaces
  • Stores up to 1,000 results
  • Integrated barcode reader (optional)